Monitoring acid-base, precipitation, complexation and redox titrations by a capacitively coupled contactless conductivity detector

文献类型: 外文期刊

第一作者: Zhang, Xuzhi

作者: Zhang, Xuzhi;Huang, Mengshi;Zhao, Jun;Qu, Keming;Huang, Mengshi;Liu, Jingquan;Yang, Wenrong

作者机构:

关键词: Micro-titration; Capacitively coupled contactless conductivity detector; Endpoint detection; Multifunctional platform; Simple titration

期刊名称:MEASUREMENT ( 影响因子:3.927; 五年影响因子:3.778 )

ISSN: 0263-2241

年卷期: 2018 年 116 卷

页码:

收录情况: SCI

摘要: Herein, an approach for versatile titration by monitoring the chemical reaction process with a developed capacitively coupled contactless conductivity detector ((CD)-D-4) was developed. A disposable glass reaction cell, in which sub-milliliter titrand was loaded, was inserted into the working head stage of the (CD)-D-4. Upon addition of titrant, the ionic strength and/or mobility of components in the solution changed, causing consequently change of conductivity. The (CD)-D-4 monitored the change in real time. A V-shaped titration curve was gained, allowing an easy identification of the endpoint. Between the elapsed time and the initial concentration of titrand, linear relationships were found for classical acid-base, precipitation, complexation and redox reaction over certain ranges, enabling the establishment of quantitative determinations. The new approach had a few charming advantages (e.g. simplicity and cost effectiveness) over contact electrochemical methods because that the solution was isolated from the working electrodes. Therefore, a development of versatile and multifunctional platform for micro-titrations is promising.