您好,欢迎访问北京市农林科学院 机构知识库!

Online Measurement of Water COD-A Comparison between Ultraviolet and Near Infrared Spectroscopies

文献类型: 外文期刊

作者: Liu Fei 1 ; Dong Da-ming 1 ; Zhao Xian-de 1 ; Zheng Pei-chao 2 ;

作者机构: 1.Beijing Acad Agr & Forestry Sci, Beijing Res Ctr Intelligent Equipment Agr, Beijing 100097, Peoples R China

2.Chongqing Univ Posts & Telecommun, Chongqing 400065, Peoples R China

关键词: COD;Ultraviolet absorption spectrum;Near infrared transmission spectra;Prediction model

期刊名称:SPECTROSCOPY AND SPECTRAL ANALYSIS ( 影响因子:0.589; 五年影响因子:0.504 )

ISSN: 1000-0593

年卷期: 2017 年 37 卷 9 期

页码:

收录情况: SCI

摘要: The spectroscopy sensing technology of water COD is an important development direction of modern environmental monitoring. Compared with traditional analytical methods, spectroscopy has more obvious advantages, such as continuous monitoring, online monitoring and fast testing, which is suitable for fixed-point and real-time monitoring of environmental water samples for COD. In this study, the ultraviolet absorption spectrum and the near infrared spectrum of real water samples were collected respectively by ultraviolet absorption spectrometry and near infrared transmission method. The COD prediction model was established by utilizing different spectral pretreatment methods combined with partial least squares regression(PLS) and multiple linear regression(MLR), and then the quantitative prediction and model parameters of ultraviolet and near infrared spectra measurement for COD were analyzed, finding that the Savitzky-Golay (SG) smoothing partial least-squares model had good prediction. Through comparison, the determination coefficients of prediction were 0. 992 1 and 0. 987 7, respectively, and RMSEP were 10. 438 6 and 5. 972 0, respectively. Ultraviolet and Near-infrared spectroscopy combined with MLR analysis model had poor prediction, with the determination coefficients of prediction 0. 928 0 and 0. 957 3, respectively. Through a comprehensive analysis of the experimental results, ultraviolet absorption spectrum prediction model in 280 '310 nm spectral region had a good performance. Near infrared spectral spectrum model had the best performance in 7 250-6 870 cm(-1) spectral region. Ultraviolet spectrum corresponding to the decision of prediction model was higher, but the near spectrum model had better stability and repeatability. Studies show that the spectrum sensing technology can be used in the quantitative predicted analysis of COD in actual water. The conclusion from the paper laid a theoretical basis for the development of portable water testing equipments.

  • 相关文献

[1]Chemometrics Method for Real-Time Measurement of Water COD Based on Laser-Induced Breakdown Spectroscopy. Ye Song,Gu Ya-hui,Zhang Wen-tao,Wang Jie-jun,Wang Xin-qiang,Dong Da-ming,Gu Ya-hui,Du Xiao-fan,Dong Da-ming. 2017

[2]A Review on Optical Measurement Method of Chemical Oxygen Demand in Water Bodies. Liu, Fei,Zheng, Peichao,Liu, Fei,Huang, Baichuan,Zhao, Xiande,Jiao, Leizi,Dong, Daming. 2016

[3]Study on Effect of Soy sauce wastewater by SBR process. Zuo, Jinlong,Wang, Xiaoyue,Yang, Xinguo,Chen, Daxiang,Wang, Xuming. 2016

[4]Study on NO2--N accumulationof soybean wastewater treatment by SBR process. Zuo, Jinlong,Chen, Daxiang,Yang, Xinguo,Wang, Xiaoyue,Wang, Xuewei,Wang, Xuming. 2016

[5]Removal of COD from piggery wastewater using a lab-scale UASB reactor. Yan, Fei,Zuo, Jinlong,Yan, Fei,Qiu, Tianlei,Wang, Xuming. 2012

[6]SBBR处理人工模拟猪场废水的试验研究. 闫飞,左金龙,李百慧,王旭明,仇天雷. 2012

[7]猪粪污厌氧发酵沼液SS、COD的混凝预处理效果研究. 李鹏,赵同科,张成军,刘宝存,郑云霞,范胜男. 2016

[8]水体COD的光谱学在线测量方法-紫外和近红外光谱比较分析. 刘飞,董大明,赵贤德,郑培超. 2017

[9]基于激光诱导击穿光谱和化学计量学的水体COD实时测量方法. 叶松,顾亚辉,杜晓凡,张文涛,汪杰君,王新强,董大明. 2017

[10]Study on Technology of Anaerobic- SBR Process for Sauce Wastewater Treatment. Zuo Jinlong,Liu Zhiwei,Li Biao,Tang Hanqing,Yan Fei,Wang Xuming. 2012

[11]A Review on Optical Measurement Method of Chemical Oxygen Demand in Water Bodies. Fei Liu,Peichao Zheng,Baichuan Huang,Xiande Zhao,Leizi Jiao,Daming Dong. 2016

作者其他论文 更多>>