Design and Implementation of National Standard Grade I Water Purification System Based on UV-Vis
文献类型: 外文期刊
作者: Zhang Heng-zhi 1 ; Ma Jie 1 ; Jia Wen-shen 1 ; Wang Ji-hua 2 ;
作者机构: 1.Beijing Informat Sci & Technol Univ, Beijing 100192, Peoples R China
2.China Dept Beijing Res Ctr Agr Stand & Testing, Beijing 100097, Peoples R China
3.Beijing Municipal Key Lab Agr Environm Monitoring, Beijing 100097, Peoples R China
关键词: UV-Vis; Grade I water; Purification system; Multiple parameter
期刊名称:SPECTROSCOPY AND SPECTRAL ANALYSIS ( 影响因子:0.589; 五年影响因子:0.504 )
ISSN: 1000-0593
年卷期: 2019 年 39 卷 8 期
页码:
收录情况: SCI
摘要: The Grade I water specified in national standards is the water commonly used in analytical laboratory. In order to analyze the quality of pure water quickly and accurately to provide scientific basis for the automation and intelligence of the equipment in analytical laboratory, this paper designed a national standard Grade I water purification system (NSGI-WPS) based on ultraviolet-visible spectroscopy (UV-VIS) by full-band (200 similar to 900 nm) UV-VIS analysis technology. The system collects modules with Raspberry Pi as the core controller and spectral detector and conductivity sensor as the sensors. As a method for the determination of soluble silicon, the improved silicon-molybdenum blue spectrophotometry realized the simultaneous on-line detection of the 254 nm absorbency, conductivity and the content of soluble silicon of the water used in analytical laboratory. In order to eliminate the impacts of noise interference on the determination test on the content of trace silicon, the system adopted Savitzky-Golay smoothing de-noising method to pretreat the spectrum. 80 groups of spectral data after smoothing were obtained via the different combinations of window width and the degree of polynomial and received unitary linear recursive analysis with the concentration of silicon standard solution 0. 004, 0. 006, 0. 008, 0. 010 and 0. 012 mg . L-1 respectively, after which the correlation spectrum between single wavelength absorbency and concentration of silicon standard solution was obtained. The results showed that when the window width is 17 and the degree of polynomial is 2, the characteristic peak of the correlation spectrum is the widest; the interval of the characteristic peak is 796 similar to 824 nm; and the wavelength of the characteristic peak is consistent with that of the peak of absorption spectra of chromogenic solutions after smoothing. It was found by comparing the distribution of the absorption spectrum of chromogenic solution of different standard silicon solutions that the concentration of silicon standard solution is positively correlated to absorption spectra. Therefore, 812. 638 nm was selected as the optimal characteristic wavelength in the test. In order to establish the relationship model between the content of soluble silicon and the absorbency of chromogenic solution, the working curve of the model is drawn with the concentration of the silicon standard solution added as the x-coordinate and the absorbency of the chromogenic solution at 812. 638 nm as the y-coordinate. The determination coefficient of the curve R-2 =0. 999 6, indicating that the model has strong fitting ability. In addition, the system management software written for NSGI-WPS realized the real-time processing and automatic control of parameters. The pure water not up to the specifications on National Standard Grade I Water used in analytical laboratory was strictly controlled by means of reverse osmosis (RO), continuous electrode ionization (EDI), mixed bed ion exchange resin and UV photooxidation and other techniques. By comparing and analyzing the changes of various parameters of tap water, grade III water and grade II water before and after purification, it was found that the values of all parameters decreased significantly after purification, among which, the maximum electrical conductivity decreased by 99. 94%, and the relative average deviation of all parameters was less than 2%. The test results showed that NSGI-WPS based on UV-Vis has such advantages as strong purification capacity, high accuracy and good robustness, and the purified water after detection, analysis and purification meets the requirements on National Standard Grade I Water used in analytical laboratory. This paper made an exploratory study on the application of UV-Vis analysis in pure water purification system.
- 相关文献
作者其他论文 更多>>
-
Study on the Predication Modeling of COD for Water Based on UV-VIS Spectroscopy and CNN Algorithm of Deep Learning
作者:Jia Wen-shen;Liang Gang;Wang Ji-hua;Jia Wen-shen;Zhang Heng-zhi;Ma Jie;Liu Xin;Jia Wen-shen;Liang Gang;Wang Ji-hua;Jia Wen-shen;Liang Gang;Wang Ji-hua
关键词:Ultraviolet visible spectrum; Convolution neural network; Chemical oxygen demand; Prediction model
-
Global sensitivity analysis of wheat grain yield and quality and the related process variables from the DSSAT-CERES model based on the extended Fourier Amplitude Sensitivity Test method
作者:Li Zhen-hai;Li Zhen-hai;Xu Xin-gang;Jin Xiu-liang;Liu Hai-long;Wang Ji-hua
关键词:global sensitivity analysis; DSSAT; EFAST; wheat; yield; grain protein content
-
An X-Ray Fluorescence Spectroscopy Pretreatment Method for Detection of Heavy Metal Content in Soil
作者:Ren Dong;Shen Jun;Ren Shun;Wang Ji-hua;Ren Dong;Shen Jun;Ren Shun;Wang Ji-hua;Wang Ji-hua;Lu An-xiang
关键词:XRF; Pretreatment; Soil heavy metal; Least square regression; Transform forward interval partial least squares
-
Research on Rapid and Non-Destructive Identification of Aging Wheat Based on ATR-Terahertz Spectroscopy Combined with PLS-DA
作者:Wang Dong;Pan Li-gang;Li An;Jin Xin-xin;Ma Zhi-hong;Wang Ji-hua;Wang Dong;Pan Li-gang;Li An;Jin Xin-xin;Ma Zhi-hong;Wang Ji-hua;Liu Long-hai;Jiang, Justin
关键词:Terahertz spectroscopy;Attenuated total reflection;Wheat;Composite score;Discriminant analysis
-
The Characteristic Spectral Selection Method Based on Forward and Backward Interval Partial Least Squares
作者:Qu Fang-fang;Ren Dong;Hou Jin-jian;Zhang Zhong;Wang Ji-hua;Hou Jin-jian;Lu An-xiang;Wang Ji-hua;Xu Hong-lei
关键词:Near-Infrared Spectroscopy;FiPLS;BiPLS;FB-iPLS;Greedy search;Characteristic intervals
-
Intercomparison of the different fusion methods for generating high spatial-temporal resolution data
作者:Shi Yue-Chan;Wang Jin-Di;Shi Yue-Chan;Yang Gui-Jun;Song Jian;Li Xin-Chuan;Wang Ji-Hua
关键词:multi-source remote sensing;fusion data;high spatial and temporal resolution;decomposition of mixed pixels
-
Establishment and Improvement of Portable X-Ray Fluorescence Spectrometer Detection Model Based on Wavelet Transform
作者:Li Fang;Wang Ji-hua;Li Fang;Wang Ji-hua;Lu An-xiang;Han Ping
关键词:X-ray fluorescence spectrometer;Heavy metal;Wavelet threshold noise filtering