摘要: Vancomycin is a glycopeptide antibiotic used to treat infections caused by Gram-positive bacteria. Due to the narrow therapeutic index of vancomycin, it is necessary to develop a sensitive and reliable analytical method to monitor the drug concentration in plasma. A novel method based on surface molecularly imprinted solid-phase extraction combined with liquid chromatography-tandem mass spectrometry for the determination of vancomycin in plasma sample was developed. The plasma sample was cleaned up through the solid-phase extraction process before the analysis. The calibration standard of vancomycin in plasma ranged between 1 and 100 ng/mL, and the correlation coefficient (r) was 0.9993. The average recoveries were from 94.3 to 104.0%, and the precision was less than 10.5%. The limit of detection and limit of quantification were 0.5 ng/mL and 1 ng/mL, respectively. The method validated was successfully used for the detection of vancomycin in mice after oral administration.